3D Linear and Volumetric Wear Measurement on Artificial Hip Joints - Validation of a New Methodology

First Published: 20/05/2021  |  Form Metrology
3d-linear-volumetric-wear-measurement-artificial-hip-joints

Mike Tukea, Andy Taylora, Anne Roquesa - Finsbury (Development), United Kingdom

Christian Maulb - RedLux Ltd, United Kingdom

Abstract

Artificial hip replacement joints have been in use to treat osteoarthritis and rheumatoid arthritis for decades. These devices consist of a metallic/ceramic ball articulating against a polyethylene or metallic/ceramic hemi-spherical cup. However, these artificial joints do wear and their clinical performance starts declining after 10 to 20 years of successful use. In the case of hard-on-hard bearings, the wear depth and volume remain very small and their measurement has remained a challenge to current technologies.

A novel device, the non-contact RedLux Artificial Hip Profiler, which is based on the chromatically encoded confocal measurement method, has been developed so that these joints can be measured accurately, reproducibly and in a short timescale. The system outputs the component radius for the unworn part, the linear depth of wear and also provides the wear volume.

The design principles are presented and the capability of the equipment to measure worn parts, with linear wear between a few microns and tens of microns is hereby characterised, and correlated with laser micrometer diameter measurements, roundness measurements of linear wear and gravimetric wear assessment.

The Artificial Hip Profiler is capable of measuring spherical bearing surfaces with a resolution of 20 nm. The correlation to benchmark methods for the three parameters considered is very high, with correlation coefficients of 1.02 and 0.95 for the linear and volumetric wear respectively. The R values were 0.988 and 0.996 respectively.

Keywords: Artificial hip joint, linear wear, volumetric wear, optical CMM

* A version of this article was published in Precision Engineering 34 (Oct 2010), pp 777 – 783

1. Introduction

An artificial hip joint is a ball-and-socket joint that can replace the arthritic human hip. The spherical head of the femur (thigh bone) moves inside the hemispherical socket of the pelvis. This movement at a rate of typically 1 – 2 million cycles per year leads to wear within the articulation 1.

When replacing a worn hip joint, the most likely surgery would involve a femoral stem being inserted into the femur (thigh bone), and a modular head being placed onto the stem via a Morse taper connection 2,3. A hemispherical cup is fitted into the reamed pelvis bone, so that the head can articulate inside the cup and replicate the standard kinematics of the joint (Figure 1 left).

Various material combinations have been used, in an effort to reduce wear at the articulating surfaces and to improve the longevity of the replaced joint 4,5. Metal or ceramic balls articulating against ultra-high molecular weight polyethylene cups with ball diameters of less than 32 mm have been the combination of choice since the 1970s 6 .

With younger patients undergoing joint replacements, ceramic-on-ceramic bearings with ball diameter of less than 32mm have recently gained in popularity 7-9. Both types of device use a relatively small diameter ball compared to the original, human joint, due to the need to reduce sliding distances in the former case, and due to manufacturing limitations in the latter. Both types of device are associated with a dislocation risk, more particularly in the young active patient 10-12.

Metal-on-metal devices have also been used historically with diameters up to 42 mm. This type of device was re-introduced in the 1990s with diameters up to 58 mm to better mimic the patient joint and reduce dislocation risks, and with the bone preserving option of a resurfacing device (Figure 1 centre, and right)  5,13,14.

image1 image3 image2-1
Figure 1: ADEPT total hip replacement (left), hip resurfacing joint (centre) and post operative X-ray of a replaced joint (right)

Successful historical metal-on-metal devices, such as the Ring or McKee Farrar devices, have been reported to achieve linear wear rates as low as 2-10 µm per year 15-18, after up to 25 years of use. Some current metal-on-metal artificial hip joints may have a similar life expectancy as they were designed to emulate these successful devices, although some concerns remain about systemic effect of elevated metal ion levels 14,19-23.

Key to an improvement in the longevity of a bearing is the understanding of its failure. With linear wear rates as low as 2 microns per year 24, this in turns relies on a method of qualifying and quantifying the wear generated on these devices.

The present paper describes a new methodology for 3D volumetric wear measurement, the RedLux Artificial Hip Profiler, and compares the measurement results to established techniques, such as gravimetric and roundness measurement.

2. Hip Joint Metrology - Background

In order to simulate wear on hip replacement devices, gait simulators have been developed and wear testing is commonly performed to standard ISO 14242. These simulate up to 10 million cycles of gait activity, and the wear is measured gravimetrically24-27. However, in these simulators, everyday activities other than gait (such as sitting on or rising from a chair or going up and down stairs), and any other, less frequent activities (such as sports) are mostly ignored. The under-estimation of wear by simulators is all the more relevant to younger patients where 10 millions cycles of load may only represent a few years of life 1.

A more accurate method of assessing wear of bearing devices is to assess the components after they have been used in service. Metal-on-metal retrievals for first and second generation devices have been analyzed for wear using coordinate measurement machines (CMM)15 or roundness machines17. The use of non-contact techniques such as laser profilometry and white light interferometry techniques have also been attempted, but with less success due to the space limitation and requirement for data stitching. The most common techniques currently used to assess hip joint bearings wear are reviewed below.

2.1 Gravimetric Method

As the hip joint wears, material is eroded from the surface of the ball and cup. This loss of material can be measured by weighing the components before and after the wear process. This alone presents a technical challenge for larger diameter low wearing metallic devices, as the gravimetric changes through wear for these devices are in the order of only a few milligrams for components weighing typically 200 g28-30. To detect these changes reliably is difficult, even for high performance scales.

Indeed, because of possible contamination from the fluids surrounding the joint and the extremely low wear rate of hard bearings, gravimetric methods generally show an increase in mass in the early wear period, before showing a decrease in mass. Correction factors have to be applied to compensate for this error31. The gravimetric method cannot give information on the shape of the wear patch and can only be used on joints that are being evaluated in wear simulators. Implanted components are not suitable for gravimetric measurement, because the exact original weight is unknown and bone ingrowth as well as protein deposits can distort the measurement significantly.

2.2 Roundness Measurement

A better way of measuring wear uses a roundness instrument to measure deviations in roundness. This is a two dimensional method that gives the user information from a trace taken around the surface.

Roundness machines can measure deviations from a circle with a resolution of typically 10 nm, and the original unworn part of the trace is compared with the worn ones. Thus the linear wear of clinical retrievals can be evaluated17,24. The roundness instrument is well suited to measuring linear wear, provided enough measurements in various planes around the joint are made and the deepest point of the wear scar is found. Kanada32 calculated that for a ball of up to 38 mm (1.5”) in diameter an average of 9 measurements are required, while for a ball above 50 mm (2”) in diameter, 18 or more measurements may be necessary to fully characterise its roundness. 3D reconstruction of the surface and the estimation of volume wear requires further measurements and assumptions and can be very time consuming17.

2.3 Coordinate Measuring Machines (CMMs)

Coordinate measuring machines (CMMs) have also been used to measure wear on artificial hip joints15,25,33. A coordinate measuring machine gives a three dimensional representation of the joint by scanning the surface with a stylus. The CMM can scan the whole spherical area of a hip joint, but it has a lower resolution, when compared to a roundness instrument, and is considerably slower as about 400-4000 points are typically necessary 15,33.

2.4 The RedLux Artificial Hip Profiler

To overcome the limitations of the techniques described above, a new form measuring instrument has been developed. The RedLux Artificial Hip Profiler (RedLux Ltd, Southampton, UK) combines the high resolution of the roundness instrument with the high coverage of the CMM, while using an automated, non-contact sensor for increased speed of measurement.

It is capable of scanning the surface of spherical objects. Unlike other optical instruments, it can scan the whole bearing surface of an artificial hip joint in a single measurement taking only a few minutes. This compares to CMM measurements that take a minimum of 20 minutes for high accuracy measurements (~4000 points) 34,35. The resulting cloud of points measured by the AHP is the raw data, and describes the form of the sphere in 3D with a resolution of 20 nm.

The instrument is a 4-axis optical CMM, as shown in Figure 2. It consists of two linear and two rotary axes, all of which utilise air bearings to achieve superior accuracy of motion. The two rotary stages carry the head or cup and the linear axes carry the sensor. All axes utilise optical encoders and linear or brushless motors.

The sensor was chosen to allow the measurement of artificial heads as well as cups. It is a point sensor, based on the chromatically encoded confocal measurement36,  which has previously been successfully applied in the semiconductor, automotive and optical industries among others37,38.

image5

Figure 2: RedLux Artificial Hip Profiler

With this sensor, the lens error commonly known as chromatic aberration is used to measure the distance to an object. The axial position of the focal point of an uncorrected lens depends on the colour (wavelength) of the light to be focussed. In the visible spectral region, the focal distance for blue light is shortest while it is longest for red light. The focal points of the other colours are located in between. Depending on the distance of the target from the focusing lens, light of just a very small wavelength region λ1 is focused on the target’s surface (Figure 3).

image4

Figure 3: Chromatically encoded confocal measurement method (© Precitec Optronik GmbH)

The focusing lens is also used to receive the backscattered light from the target’s surface and to focus it into an optical fibre and later a spectrometer. The wavelength scale of the spectrometer is calibrated versus the distance to the scattering surface, giving the required output signal.

By combining the sensor signal with the knowledge over the exact position of all 4 stages, a 3D representation of the surface can be created in PC memory. Once these data are available, they can be processed and analysed. For a typical 50 mm diameter head, 40,000 points are measured, while for the corresponding cup 30,000 points are taken. For detailed investigations, over one million data points are available and a measurement all the way to the edge of the cup is possible. 

3. Materials and Methods

3.1 Accuracy of Measurement – Calibration Ball

The underlying accuracy of the RedLux Artificial Hip Profiler was assessed against a calibrated reference ball (Table 1).

Accuracy Grade 3 (< 80 nm)
Diameter [mm] 24.99985
Roundness [μm] 0.04
Material Tungsten Carbide
Traceability To National Standards via Gauge Blocks serial no: 10054
Table 1: Specifications of calibrated reference ball

The form error on this ball is negligible for the purpose of machine error assessment and any measurement of this ball will show the machine error. The calibration ball has been measured to assess the errors for sphericity and radius measurements.

The thermal influence on radius measurements was assessed by repeatedly measuring the calibration ball over a two day period, while ambient temperature varied.

3.2 Wear Assessment – Simulated Wear

The current standard for volumetric wear assessment on simulator samples is the gravimetric method as introduced above. To compare the Artificial Hip Profiler with the gravimetric method, a test was set up whereby 6 fully finished modular heads were worn and the wear assessed with both techniques. Artificially worn parts were preferred to parts worn in a wear simulator as the latter can easily be contaminated by fluids surrounding the joint, making an accurate weight measurement difficult. Six intact ADEPT (Finsbury Orthopaedics Ltd, Leatherhead, UK) modular head components (42 mm diameter) were measured as detailed below before proceeding. Worn patches were created by polishing the surface with a polishing mop at a given, known position for 1 to 5 minutes using a hand held powered polishing tool.

The procedure below was followed ensuring that traceability was kept at all times.

  • Measure diameter on a laser micrometer LSM-506 and LSM-6000 (Mitutoyo).
  • Measure weight with Sartorius ME235S balance, resolution 0.01 mg, linearity ≤ 0.1 mg following standard hip simulator cleaning procedure.
  • Create wear patches and clean.
  • Measure wear on Gravimetric Measurement with Sartorius ME235S balance.
  • Measure on roundness machine (series 211, Mitutoyo roundness RA400, Kawasaki, Japan), by measuring equatorial traces at 4 to 5 positions along the head axis, spaced approximately 2.5 mm, at the position of the wear patch. Each trace is scanned and loaded in a CAD package (SolidWorks 2009 Sp04), and a circle is fitted to the unworn part of the trace; the maximum linear wear depth is measured as the maximum deviation from this circle in the radial direction, in 2D. As each trace is taken in an equatorial plane, the equivalent maximum linear wear value in the sphere radius direction (3D) is derived from simple trigonometry.
  • Measure with Artificial Hip Profiler (after wear only). Output the diameter of the unworn part of the sphere, linear wear and volumetric wear (assuming a cobalt chrome density of 8.33 g/cm3).

All measurements were repeated three times for each modular head.

3.3 Artificial Hip Profiler Post-Processing Method

With the Artificial Hip Profiler, the data requires post processing to obtain the volumetric wear from the raw data, as illustrated in Figure 4.

In a first step, a sphere is fitted to the raw data set by linear least-squares sphere fit (a). The result is slightly skewed, as the wear scar impacts the sphere fit. The wear scar is therefore excluded from processing (b, c), and only the original, unworn surface is used to compute the reference sphere. Finally, the previously excluded data points are re-included in the display, showing the wear scar in relation to the unworn bearing surface (d).

image7-1 image6

a) Raw data

b) Wear scar marked

image9-1 image8
c) Wear scar excluded d) Wear scar relative to unworn surface
Figure 4: Wear measurement – illustration of sphere fit to unworn part of a head

The wear volume is the difference between the best-fit sphere and the measured data set.

4. Results and Discussion

4.1 Accuracy of Measurement – Calibration Ball

Two measurements were performed on the calibration ball; firstly, a measurement of sphericity and secondly, a measurement of radius over time. This was to establish the base accuracy of measurement.

Figure 5 shows the result of a sphericity measurement on the calibration ball. 95% of the measured points are within 146 nm of the best-fit sphere, showing the measurement error of the system. To achieve this level of accuracy, the system is located on a vibration isolation table and stage errors are compensated for numerically. Temperature drift during measurement is negligible, due to the short measurement time.

image12-1

Figure 5: Sphericity of Calibration Ball

To establish the instrument performance versus external temperature, 866 repeat measurements of the calibration ball over a period of 2 days were performed. On the left axis in Figure 6 the measured radius is shown as the red curve, while the right axis shows the ambient temperature with the curve in yellow. The temperature varied by ± 2.1 ºC, while the radius varied by ± 1.6 μm over the same 2 day period. It should be noted that the thermal expansion of the tungsten carbide ball was not taken into account. It is expected that this would have resulted in a slight improvement of the measurement results.

image10-1

Figure 6: Radius repeatability measurements.

Temperature effects are numerically compensated for and only have a small influence on the measurement of component radius, even in an uncontrolled environment. The influence of temperature on the measurement of sphericity and wear is negligible, because these are measured relative to the best fit sphere. If the radius of a 25 mm head changes by 1 μm due to a temperature change, the linear wear measurement of a wear scar of say 25 μm changes by

25 μm x 1 μm / 25 mm = 1 nm.

On a wear patch of 500 mm2 this causes an error of 0.0005 mm3 and can be neglected.

4.2 Wear Assessment – Simulated Wear

The following table summarises the results obtained with the benchmark methods and the Artificial Hip Profiler (AHP) for the 6 heads under study, measured three times over two days in a non-temperature controlled environment (average and standard deviation of three samples shown).

Table 2: Comparison between standard measurement methods and the AHP (average (stdev) of three measurements)

Component Id

AHP Measurement time (min and sec)

Diameter (laser micrometer) mm

Diameter (AHP) mm

Linear Wear
(Roundness) μm

Linear Wear
(AHP) μm

Volume Loss
(gravimetric) mm3

Volume Loss
(AHP) mm3

1

1’53-2’07

41.8484
(0.0024)

41.849089 (0.00083)

10.3 (0.3)

10.7 (0.3)

0.313 (0.006)

0.280 (0.022)

2

1’51-2’06

41.8499
(0.0019)

41.850737 (0.00251)

23.3 (0.4)

22.9 (0.3)

0.600 (0.004)

0.575 (0.016)

3

1’52-2’06

41.8477
(0.0013)

41.849245 (0.00193)

2.0 (0.1)

2.6 (0.1)

0.069 (0.004)

0.047 (0.005)

4

1’51-2’07

41.8478
(0.0013)

41.847964  (0.00036)

9.3 (0.2)

11.0 (0.1)

0.306 (0.005)

0.293 (0.001)

5

1’52-2’07

41.8518
(0.0023)

41.852021  (0.00153)

11.7 (0.3)

12.2 (0.2)

0.404 (0.002)

0.392 (0.012)

6

1’53-2’07

41.8453
(0.0023)

41.846755  (0.00144)

3.6 (0.1)

3.9 (0.2)

0.067 (0.006)

0.064 (0.003)

The diameter measurements are in good agreement between laser micrometer and Artificial Hip Profiler, with the variation likely to be caused by temperature variations of the components.

Linear wear measurements are again in very good agreement, with the measurements by the Artificial Hip Profiler being generally larger than those measured with the roundness instrument. The deepest part of the wear scar on these components extends over only one or two millimetres and with the roundness instrument it is difficult to find the exact location of the valley.

Thirdly, the Artificial Hip Profiler provides comparable results to the gravimetric method. The maximum difference for measured volume loss is 0.033 mm3 Potential sources of errors arise from the volume estimation used in the Artificial Hip Profiler. Despite these, there is a strong agreement between the two techniques, with the AHP providing the advantage of a quick measurement without the need for extensive cleaning.

The correlations between the optical and control methods for the measurement of wear are shown in Figure 7. The correlation coefficients were 1.02 and 0.95 for the linear and volumetric wear respectively. The R values were 0.988 and 0.996 respectively. The Bland-Altman plots are shown in Figure 8.

image11-1 image14-1
Figure 7: Correlation curves – linear wear correlation with Mitutoyo roundness and volumetric wear correlation with gravimetric method.

image15-1

image16-1

Figure 8: Bland-Altman plot comparing maximum linear wear and volumetric wear obtained using the AHP and conventional methods

It is expected that these results can be improved upon, when measurements of component form are performed prior to wear and a before/after comparison is carried out. This work is currently under way and subject to a future publication.

The new device can perform the measurement of diameter, linear wear and volumetric wear in a single measurement taking only a few minutes.

In contrast to the gravimetric method, the Artificial Hip Profiler eliminates the need for extensive cleaning of components. It provides information on the wear volume as well as the shape and location of the wear patch and works on both wear simulator samples and retrievals.

In comparison to a roundness instrument, the Artificial Hip Profiler provides a 3D scan of the surface, permitting volumetric wear measurement and geometric information on the wear scar.

The results show a strong correlation to the results obtained using conventional methods.

5. Conclusions

The Artificial Hip Profiler, based on the chromatically encoded confocal measurement method, had been developed to measure spherical bearing surfaces for total joint replacement with diameter ranging from 28 to 58mm in diameter. It is capable of scanning the parts in a few minutes.

On worn parts, a 3D visualisation of the surface can be obtained, and wear patches of depth as small as a few microns and as high as a few hundred microns can be analysed. The correlation of the results obtained in terms of unworn sphere diameter, linear wear depth and volume wear loss estimation all correlate extremely well with results obtained using benchmark methods, with correlation coefficients of 0.95 and 1.02 and R ratios of 0.986 and 0.996.

Acknowledgements

The authors would like to thank Dr Bill Walter and Christina Esposito, University of New South Wales (Australia), and Dr Richie Gill, University of Oxford (United Kingdom), for valuable discussions during the preparation of this paper.

Reference List

  1. Silva M, Shepherd E, Jackson W, Dorey F, Schmalzried T: Average patient walking activity approaches 2 million cycles per year: pedometers under-record walking activity. Journal of Arthrosplasty. 2002;17:693-697.
  2. Faulkner A, Kennedy L, Baxter K, Donovan J, Wilkinson M, Bevan G. Effectiveness of hip prostheses in primary total hip replacement: a critical review of evidence and an economic model. http://www.hta.ac.uk/fullmono/mon206.pdf. Health Technology Assessment, 2(6)[NHS R&D HTA Programme]. 1998. 10-5-2010. 
  3. Laupacis A, Bourne R, Rorabeck C, Feeny D, Wong C, Tugwell P et al.: The effect of elective total hip replacement on health-related quality of life. J Bone Joint Surg Am. 1993;75:1619-1626.
  4. Schmalzried TP, Callaghan JJ: Wear in total hip and knee replacements. J Bone Joint Surg Am. 1999;81:115-136.
  5. Heisel C, Silva M, Schmalzried TP: Bearing surface options for total hip replacement in young patients. Instr Course Lect. 2004;53:49-65.
  6. Ratner B, Hoffman A, Schoen F, Lemons J: Biomaterials Science - An introduction to Materials in Medicine, 1st edition edn. Academic Press; 1996.
  7. Bizot P, Banallec L, Sedel L, Nizard R: Alumina-on-alumina total hip prostheses in patients 40 years of age or younger. Clin Orthop Relat Res. 2000;68-76.
  8. Dudkiewicz I, Salai M, Israeli A, Amit Y, Chechick A: Total hip arthroplasty in patients younger than 30 years of age. Isr Med Assoc J. 2003;5:709-712.
  9. Garcia-Rey E, Cruz-Pardos A, Garcia-Cimbrelo E: Alumina-on-alumina total hip arthroplasty in young patients: diagnosis is more important than age. Clin Orthop Relat Res. 2009;467:2281-2289.
  10. Kim YH, Choi Y, Kim JS: Influence of patient-, design-, and surgery-related factors on rate of dislocation after primary cementless total hip arthroplasty. Journal of Arthrosplasty. 2009;24:1258-1263.
  11. Skeels MD, Berend KR, Lombardi AV, Jr.: The dislocator, early and late: the role of large heads. Orthopedics. 2009;32.
  12. Sexton SA, Walter WL, Jackson MP, De Steiger R, Stanford T: Ceramic-on-ceramic bearing surface and risk of revision due to dislocation after primary total hip replacement. J Bone Joint Surg Br. 2009;91:1448-1453.
  13. Amstutz HC, Le Duff MJ: Background of metal-on-metal resurfacing. Proc Inst Mech Eng H. 2006;220:85-94.
  14. Hummel MT, Malkani AL, Yakkanti MR, Baker DL: Decreased dislocation after revision total hip arthroplasty using larger femoral head size and posterior capsular repair. Journal of Arthrosplasty. 2009;24:73-76.
  15. McKellop H, Park S, Chiesa R: In Vivo Wear of three types of metal on metal Hip Prothesis during 2 decades of use. Clinical Orthopaedics and Related Research. 1996;329S:128-140.
  16. Kothari M, Bartel DL, Booker JF: Surface geometry of retrieved McKee-Farrar total hip replacements. Clin Orthop Relat Res. 1996;S141-S147.
  17. Willert HG, Buchhorn GH, Gobel D, Koster G, Schaffner S, Schenk R et al.: Wear behavior and histopathology of classic cemented metal on metal hip endoprostheses. Clin Orthop Relat Res. 1996;S160-S186.
  18. Sauve P, Mountney J, Khan T, De Beer J, Higgins B, Grover M: Metal ion levels after metal-on-metal Ring total hip replacement: A 30-YEAR FOLLOW-UP STUDY. J Bone Joint Surg Br. 2007;89-B:586-590.
  19. Quesada MJ, Marker DR, Mont MA: Metal-on-metal hip resurfacing: advantages and disadvantages. Journal of Arthrosplasty. 2008;23:69-73.
  20. De Haan R, Campbell PA, Su EP, De Smet KA: Revision of metal-on-metal resurfacing arthroplasty of the hip: the influence of malpositioning of the components. J Bone Joint Surg Br. 2008;90:1158-1163.
  21. Nunley RM, Della Valle CJ, Barrack RL: Is patient selection important for hip resurfacing? Clin Orthop Relat Res. 2009;467:56-65.
  22. Della Valle CJ, Nunley RM, Raterman SJ, Barrack RL: Initial American experience with hip resurfacing following FDA approval. Clin Orthop Relat Res. 2009;467:72-78.
  23. Fowble VA, dela Rosa MA, Schmalzried TP: A comparison of total hip resurfacing and total hip arthroplasty - patients and outcomes. Bull NYU Hosp Jt Dis. 2009;67:108-112.
  24. Tuke MA, Scott G, Roques A, Hu XQ, Taylor A: Design considerations and life prediction of metal-on-metal bearings: the effect of clearance. J Bone Joint Surg Am. 2008;90 Suppl 3:134-141.
  25. Affatato S, Spinelli M, Zavalloni M, Mazzega-Fabbro C, Viceconti M: Tribology and total hip joint replacement: current concepts in mechanical simulation. Med Eng Phys. 2008;30:1305-1317.
  26. Scholes SC, Unsworth A: Comparison of friction and lubrication of different hip prostheses. Proc Inst Mech Eng H. 2000;214:49-57.
  27. Vassiliou K, Scholes SC, Unsworth A: Laboratory studies on the tribology of hard bearing hip prostheses: ceramic on ceramic and metal on metal. Proc Inst Mech Eng H. 2007;221:11-20.
  28. Anissian HL, Stark A, Good V, Dahlstrand H, Clarke IC: The wear pattern in metal-on-metal hip prostheses. J Biomed Mater Res. 2001;58:673-678.
  29. Chan FW, Bobyn JD, Medley JB, Krygier JJ, Yue S, Tanzer M: Engineering issues and wear performance of metal on metal hip implants. Clin Orthop Relat Res. 1996;96-107.
  30. Dowson D, Hardaker C, Flett M, Isaac GH: A hip joint simulator study of the performance of metal-on-metal joints: Part II: Design. Journal of Arthrosplasty. 2004;19:124-130.
  31. Scholes SC, Unsworth A: The effects of proteins on the friction and lubrication of artificial joints. Proc Inst Mech Eng H. 2006;220:687-693.
  32. Kanada T: Evaluation of spherical form errors – Computation of sphericity by means of minimum zone method and some examinations with using simulated data. Precision Engineering. 1995;17:281-289.
  33. Schmidt M, Weber H, Schon R: Cobalt chromium molybdenum metal combination for modular hip prostheses. Clin Orthop Relat Res. 1996;S35-S47.
  34. Reinisch G, Judmann KP, Lhotka C, Lintner F, Zweymuller KA: Retrieval study of uncemented metal-metal hip prostheses revised for early loosening. Biomaterials. 2003;24:1081-1091.
  35. Joyce T, Langton D, Nargol AVN: Wear measurements from explanted Metal-on-Metal hip resurfacing prostheses. 2010; Read at the 56th Annual meeting of the Orthopaedic Research Society. Poster 2338.
  36. BATTELLE-INSTITUT E.V. Device for measuring distances between an optical element with high chromatic aberration and an object. WO/1988/010406[WO/1988/010406]. 1988. 
  37. Boltryk P, Hill M, McBride J: Comparing laser and polychromatic confocal optical displacement sensors for the 3D measurement of cylindrical artefacts containing microscopic grooved structures. Wear. 2009;266:498-501.
  38. Kunkel M, Schulze J: Noncontact measurement of central lens thickness. GLASS SCIENCE AND TECHNOLOGY. 2005;78:1-4.

Metrology for You.

We’re helping organisations all over the world reap the benefits of world-leading metrology. Speak to our experts.

Get in touch